Résistivité et Seebeck à haute température (ZEM3)

11 mars 2011

High temperature resistivity and Seebeck measurement


Description :

This equipement measure in the same time the Seebeck coefficient and the resistivity of thermoelectrics materials. The thermoelectrics properties could be obtained in a large range of materials, semiconductors, metals and ceramics included.


The use of imaging furnace allow a good temperature control. Measures are done for each temperature gradient at computer controlled temperature that allow an automatic measurement eliminating the "dark electromotive force".


Measurement atmosphere : Helium low pressure.

How does it work ?



The thermocouples A and B measure the voltage and the temperature.
Coefficient Seebeck = DV/(T°A-T°B)
Temperature = (T°A+T°B)/2
Resistance = V/I

Possibilities :

Temperature range : RT - 700°C
Measurement during heating and cooling.
Maximum of measurement point in the temperature range : 125
Sample size : high = 5 to 22 mm
width or diametre = 2 to 4 mm

Contact : Emmanuel GUILMEAU